Fixture Level
[
Pick-up Type
]-
Dual Abutment
SimpleLine ll Surgical / Prosthesis Manual
22
[
Multiple Units
]
Final prosthesis. Use positioning jig to transfer the
abutment from the model to the intraoral
and then tighten it with 25~30N‧cm.
Re-tighten it after 15 minutes.
Cement the final prosthesis and make
occlusal adjustment.
Create an access hole for
pick-up coping screw.
Completion of Wax-up. Metal framework.
SCRP- Lab Side
Final prosthesis. Use positioning jig to transfer the
abutment from the model to the intraoral
and tighten with to 25~30N‧cm.
Re-tighten it after 15 minutes.
* In the process of seating the prosthesis, the components can be rebounded by gingival tissue.
In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.
* In the process of seating the prosthesis, the components can be rebounded by gingival tissue.
In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.
Cement the final prosthesis and make
occlusal adjustment.
Chairside
SCRP- Chairside